REF: HF2POSTD HomeFill Oxygen System HomeFill cylinder D with CGA870 Compatible Post Valve Recall

REF: HF2POSTD HomeFill Oxygen System HomeFill cylinder D with CGA870 Compatible Post Valve

FDA recall Class II

Ongoing

FDA lists this Class II recall for REF: HF2POSTD HomeFill Oxygen System HomeFill cylinder D with CGA870 Compatible Post Valve. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3223-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 28, 2026
FDA report date
Sep 30, 2026
Quantity in commerce
25 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

REF: HF2POSTD HomeFill Oxygen System HomeFill cylinder D with CGA870 Compatible Post Valve

Affected model & UDI checker

Search the exact identifiers published for Z-3223-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
HF2POSTDREF: HF2POSTD HomeFill Oxygen System HomeFill cylinder D with CGA870 Compatible Post ValveCheck source noticeNot specified
Not specifiedREF: HF2POSTD HomeFill Oxygen System HomeFill cylinder D with CGA870 Compatible Post ValveCheck source notice00850018761376
FDA code information

UDI:00850018761376/ Serial #s: 25EKMA00BAC 25EKMA009L3 25EKMA009CM 25EKMA009J8 25GKMA00077 25GKMA0007B 25GKMA0009Z 25GKMA0004U 25GKMA000CS 25GKMA000CR 25GKMA000CX 25GKMA000D0 25GKMA000LL 25GKMA000N1 25GKMA000UD 25GKMA000D1 25GKMA000TN 25GKMA000D2 25GKMA000UC 25GKMA000LK 25GKMA000UA 25GKMA000UB 25GKMA000UE 25GKMA000CY 25GKMA000UF

View the original FDA recall notice

Why was this product recalled?

Labeling; Oxygen Cylinders do not contain required medical gas labeling

What actions did Ventec Life Systems, Inc. take?

On July 28, 2026, Ventec Life Systems Inc. issued a Urgent: Medical Device Recall notification to affected consignees via: certified mail. Ventec ask consignees to take the following actions: 1. Immediately discontinue use of affected HomeFill Oxygen Cylinders. 2. Inspect current stock, quarantine/segregate any unused affected products, remove affected product from clinical service and segregate to prevent further use until the enclosed label correction has been completed. 3. Use an unaffected oxygen cylinder for patient support. 4. If you have distributed or transferred affected product to other facilities, caregivers, or end users, you must promptly notify them of this recall and provide a copy of this notification. 5.. Confirm receipt of this notification letter by returning the enclosed Customer Response Form to React Health WITHIN 7 DAYS OF RECEIPT. 6. Upon receipt of the Customer Response Form, React Health will arrange for shipment of the label and instructions to apply the label to affected cylinders along with an update confirmation letter. 7. Update the affected product and return the completed confirmation letter to React Health. Contact Customer Support at 863-226 -6285 to discuss completing the form, if necessary.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology