FDA RECALL RECORD · Z-3225-2026

CombiSet Hemodialysis Blood Tubing Set with Attached Priming Set and Transducer Protectors; Recall

CombiSet Hemodialysis Blood Tubing Set with Attached Priming Set and Transducer Protectors; Model Number: 03-2522-1;

FDA recall Class II

Ongoing

FDA lists this Class II recall for CombiSet Hemodialysis Blood Tubing Set with Attached Priming Set and Transducer Protectors;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3225-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 29, 2026
FDA report date
Sep 30, 2026
Quantity in commerce
1743026 units
Distribution
See key details
FDA cause
Use error

What is being recalled?

CombiSet Hemodialysis Blood Tubing Set with Attached Priming Set and Transducer Protectors; Model Number: 03-2522-1;

Affected model & UDI checker

Search the exact identifiers published for Z-3225-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
03-2522-1CombiSet Hemodialysis Blood Tubing Set with Attached Priming Set and Transducer Protectors;Check source noticeNot specified
Not specifiedCombiSet Hemodialysis Blood Tubing Set with Attached Priming Set and Transducer Protectors;Check source notice00840861102402
FDA code information

Model Number: 03-2522-1; UDI-DI: 00840861102402 (Bag), 10840861102409 (Case); Lot codes: 26AR01010, 26BR01072, 26CR01067, 26DR01160, 26DR01198, 26AR01011, 26BR01074, 26CR01069, 26DR01161, 26DR01203, 26AR01034, 26BR01076, 26CR01070, 26DR01162, 26DR01206, 26AR01035, 26BR01077, 26CR01072, 26DR01163, 26DR01207, 26AR01046, 26BR01078, 26CR01073, 26DR01165, 26DR01209, 26AR01047, 26BR01079, 26CR01074, 26DR01166, 26DR01210, 26AR01056, 26BR01082, 26CR01075, 26DR01167, 26DR01255, 26AR01057, 26BR01083, 26CR01076, 26DR01168, 26ER01022, 26AR01058, 26BR01084, 26CR01079, 26DR01170, 26ER01023, 26AR01059, 26BR01085, 26CR01080, 26DR01171, 26ER01024, 26AR01230, 26BR01087, 26CR01081, 26DR01172, 26ER01029, 26BR01007, 26BR01088, 26CR01082, 26DR01173, 26ER01034, 26BR01035, 26BR01210, 26CR01083, 26DR01174, 26ER01035, 26BR01036, 26CR01014, 26CR01084, 26DR01175, 26ER01037, 26BR01037, 26CR01045, 26CR01085, 26DR01177, 26ER01039, 26BR01039, 26CR01046, 26CR01086, 26DR01180, 26ER01040, 26BR01045, 26CR01047, 26CR01087, 26DR01182, 26ER01042, 26BR01049, 26CR01048, 26CR01088, 26DR01183, 26ER01044, 26BR01051, 26CR01049, 26CR01090, 26DR01184, 26ER01048, 26BR01052, 26CR01051, 26CR01091, 26DR01185, 26ER01049, 26BR01053, 26CR01052, 26CR01092, 26DR01186, 26ER01054, 26BR01057, 26CR01054, 26CR01093, 26DR01187, 26ER01055, 26BR01058, 26CR01056, 26CR01094, 26DR01188, 26ER01056, 26BR01059, 26CR01057, 26CR01206, 26DR01189, 26ER01062, 26BR01060, 26CR01058, 26CR01207, 26DR01190, 26ER01065, 26BR01063, 26CR01059, 26CR01208, 26DR01191, 26ER01066, 26BR01064, 26CR01060, 26CR01209, 26DR01193, 26ER01074, 26BR01065, 26CR01061, 26CR01210, 26DR01194, 26ER01266, 26BR01066, 26CR01064, 26DR01013, 26DR01195, 26ER01268, 26BR01067, 26CR01065, 26DR01158, 26DR01196, 26ER01282, 26BR01069, 26CR01066, 26DR01159, 26DR01197;

View the original FDA recall notice

Why was this product recalled?

Communication to recommend reviewing the IFU and ensuring that all treatment preparation and handling steps are performed as instructed.

What actions did Fresenius Medical Care Renal Therapies Group, LLC take?

On July 30, 2026 IMPORTANT MEDICAL DEVICE INFORMATION letters were sent to customers. Letter provides information and instructions on the use of the device. As part of routine best practice, we recommend reviewing the IFU and ensuring that all treatment preparation and handling steps are performed as instructed. There are potential risks to patient health that are associated with blood loss, treatment delay, or air entering the system. Ensure all steps of the IFU are followed, that the Transducer Protector connection to the bloodline monitor line is tight, and that the transducer protector and female luer connections are finger tight to maintain a closed system. Please contact FMCRTG Technical Service at 1-800-227-2572 if you have any further questions. For any clinical or medical questions, please contact the Global Medical Information and Education Office at 1-855-616-2309.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.