FDA RECALL RECORD · Z-3231-2026

VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811 Recall

VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811

FDA recall Class II

Ongoing

FDA lists this Class II recall for VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3231-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jul 24, 2026
FDA report date
Sep 30, 2026
Quantity in commerce
97680 units
Distribution
See key details
FDA cause
Packaging process control

What is being recalled?

VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811

Affected model & UDI checker

Search the exact identifiers published for Z-3231-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedVITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811Check source notice03573026637712
FDA code information

UDI-DI 03573026637712; Lot Number 1333393504

View the original FDA recall notice

Why was this product recalled?

The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did BIOMERIEUX, Inc. take?

An URGENT FIELD SAFETY NOTICE notification letter was sent to customers beginning 7/24/26. Required actions Dispose of any remaining inventory VITEK 2 AST-N461 TEST KIT (Ref 424811, Lot 1333393504) that were manufactured with Poucher 8 with timestamps from 10:06:00 to 11:11:59, as identified by product labeling. Inspect each pouch in inventory and dispose of cards with "P8" and timestamps between 10:06:00 to 11:11:59 on the label. Alternatively, dispose of all cards from Lot 1333393504 remaining in inventory. In accordance with your facility s policies, you may consider performing a retrospective review of clinically significant resistant results. Among tests previously performed, identify any possible false resistant results that may have occurred, analyze the related risks, and determine appropriate actions, if applicable. Please ensure that this information is communicated to all relevant personnel within your laboratory, retain a copy in your records, and forward this information to all parties that may use this product, including those to whom you may have transferred the product. Please complete the Acknowledgement Form in Attachment A and return it to your local bioMrieux representative to confirm receipt of this notice. It is important that you return the acknowledgement form to bioMrieux even if you determine that this product correction notice does not impact your facility. As a reminder, please visit the Resource Center to confirm if a reference package insert is associated with a Field Corrective Action or Field Safety Corrective Action. https://resourcecenter.biomerieux.com/ If you require additional assistance or have any questions, please contact your local bioMrieux Customer Service representative.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology