Achieva 3.0T; Model Numbers (REF): 1. 781177, 2. 781277, 3. 781278, 4. 781344, 5. 781345; Recall
Achieva 3.0T; Model Numbers (REF): 1. 781177, 2. 781277, 3. 781278, 4. 781344, 5. 781345;
FDA recall Class II
Ongoing
FDA lists this Class II recall for Achieva 3.0T; Model Numbers (REF): 1. 781177, 2. 781277, 3. 781278, 4. 781344, 5. 781345;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3233-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 21, 2026
- FDA report date
- Sep 30, 2026
- Quantity in commerce
- 210
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
Achieva 3.0T; Model Numbers (REF): 1. 781177, 2. 781277, 3. 781278, 4. 781344, 5. 781345;
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
1. Model Numbers (REF): 781177; UDI-DI: N/A; Serial Numbers: 17144, 17376, 17448, 24049, 17099, 24138, 24060, 17607, 17079, 17146, 17189, 17379, 17469, 17270, 17432, 17279, 17233, 17295, 17446, 17362, 24068, 24096, 17139, 24037, 24099, 17535, 24079, 17030, 17043, 17056, 17114, 17149, 17161, 17167, 17170, 17179, 17216, 17217, 17235, 17251, 17316, 17319, 17365, 17453, 17494, 17523, 17558, 17561, 24038, 24070, 24089; 2. Model Numbers (REF): 781277; UDI-DI: N/A; Serial Numbers: 34289, 34213, 34276, 34113, 34123, 34239, 34264, 34274, 34285, 34095, 34138, 34305, 34118, 34106, 34159, 34241, 34263, 34139, 34002, 34029, 34034, 34071, 34201; 3. Model Numbers (REF): 781278; UDI-DI: N/A; Serial Numbers: 38326, 38346, 38043, 38098, 38108, 38152, 38157, 38194, 38210, 38212, 38215, 38231, 38233, 38236, 38238, 38240, 38267, 38269, 38270, 38271, 38284, 38312, 38321, 38328, 38332, 38338, 38339, 38340, 38344, 38349, 38367, 38370, 38401, 38402, 38405, 38407, 38408, 38412, 38415, 38420, 38421, 38423, 38425, 38426, 38427, 38430, 38431, 38432, 38433, 38437, 38439, 38441, 38442, 38443, 38445, 38447, 38448, 38450, 38453, 38454, 38456, 38457, 38458, 38288, 38079, 38139, 38228, 38145, 38034, 38251, 38278, 38207, 38061, 38086, 38095, 38115, 38133, 38017, 38064, 38112, 38143, 38160, 38183, 38229, 38256, 38296, 38297, 38315, 38331, 38272, 38322, 38323, 38337, 38345, 38077, 38200, 38268, 38090, 38188, 38093, 38012, 38099, 38003, 38057, 38072, 38085, 38016, 38033, 38091, 38103, 38173, 38341, 38232; 4. Model Numbers (REF): 781344; UDI-DI: N/A; Serial Numbers: 73005, 73008, 73009, 73006, 73010; 5. Model Numbers (REF): 781345; UDI-DI: N/A; Serial Numbers: 74027, 74035, 74039, 74047, 74052, 74001, 74010, 74026, 74046, 74005, 74007, 74021, 74030, 74040, 74044, 74048, 74032, 74037;
Why was this product recalled?
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
What actions did Philips North America Llc take?
On August 24, 2026 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: " To prevent this issue from occurring, o Use the standard 3D Vane scan protocol from the Philips Examcard Database, or o For custom 3D Vane scan protocols, use a maximum of 480 for the reconstruction matrix and use a maximum of 167 for the number of slices. " You may continue to use affected systems in accordance with their intended use. " Circulate this notice to all users or new owners of this device so that they are aware of the issues and associated hazard/harm. " Please retain this Urgent Medical Device letter with your system(s) until the software upgrade is installed; ensure the notice is in a place likely to be seen/viewed. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule time for a Field Service Engineer (FSE) to install a software upgrade to resolve the issue (reference 2026-PD-MR-006).
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3233-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology