FDA RECALL RECORD · Z-3253-2026

nGenuity Patient Monitor with Capnography (ETCO2) CAT 8100E1-ST-US, CAT 8100E1-US, CAT 8100EP1, CAT 8100EP1-ST, CAT 8100EP1-ST-US, CAT 8100EP1-US, CAT 8100EP1R Recall

nGenuity Patient Monitor with Capnography (ETCO2) CAT 8100E1-ST-US, CAT 8100E1-US, CAT 8100EP1, CAT 8100EP1-ST, CAT 8100EP1-ST-US, CAT 8100EP1-US, CAT 8100EP1R

FDA recall Class II

Ongoing

FDA lists this Class II recall for nGenuity Patient Monitor with Capnography (ETCO2) CAT 8100E1-ST-US, CAT 8100E1-US, CAT 8100EP1, CAT 8100EP1-ST, CAT 8100EP1-ST-US, CAT 8100EP1-US, CAT 8100EP1R. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3253-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 6, 2026
FDA report date
Sep 30, 2026
Quantity in commerce
1,007
Distribution
See key details
FDA cause
Process design

What is being recalled?

nGenuity Patient Monitor with Capnography (ETCO2) CAT 8100E1-ST-US, CAT 8100E1-US, CAT 8100EP1, CAT 8100EP1-ST, CAT 8100EP1-ST-US, CAT 8100EP1-US, CAT 8100EP1R

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

UDI-DI +B4388100EP1US02 All units with CO2 option SN beginning with 424, 125, 225, 325, 425, 126, 226, 326, 426, date range 2024-1st qt 2026

View the original FDA recall notice

Why was this product recalled?

Complaint trend where "Checkline Error" alarm presents at device setup, rendering the CO2 monitoring function inoperable. May result in procedure delay.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Criticare Technologies, Inc. take?

Firm posted its initial notice on May 6, 2026. A follow-up letter was sent on August 31, 2026 with additional instructions. Customers were instructed to do the following: 1. Be aware of the Checkline Error alarm. If it appears on your CO2 section at startup, follow the steps provided in the letter. 2. Verify your backup plan. Confirm that your practice has access to an alternate CO2 monitoring source. 3. Retain the letter and note your unit s serial number for your records. 4. Contact the firm if you have any questions.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.