FDA RECALL RECORD · Z-3275-2026

EnCor EnCompass Breast Biopsy and Tissue Removal System Console ( Recall

EnCor EnCompass Breast Biopsy and Tissue Removal System Console (model number: ENCPCONSOLE) and Driver (model number: ENCPDRIVER) with software

FDA recall Class II

Ongoing

FDA lists this Class II recall for EnCor EnCompass Breast Biopsy and Tissue Removal System Console (. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-3275-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Aug 20, 2026
FDA report date
Oct 7, 2026
Quantity in commerce
99
Distribution
See key details
FDA cause
Software design

What is being recalled?

EnCor EnCompass Breast Biopsy and Tissue Removal System Console (model number: ENCPCONSOLE) and Driver (model number: ENCPDRIVER) with software

Affected model & UDI checker

Search the exact identifiers published for Z-3275-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedEnCor EnCompass Breast Biopsy and Tissue Removal System Console (Check source notice00801741235221
FDA code information

Software Versions 1.1.1.10207 and 1.2.0.12133. UDI-DI: 00801741235221 and 00801741234248. Serial Numbers: VTLNPX06, VTLNPX0F, VTLNPX0D, VTLNPX08, VTLNPX0K, VTLNPX0A, VTLNPX0C, VTLNPX0L, VTLNPX0J, VTLNPX04, VTLNPX0H, VTLNPX02, VTLPPX1G, VTLPPX1D, VTLPPX1C, VTLPPX1M, VTLPPX2H, VTLPPX2I, VTLPPX2F, VTLQPX15, VTLQPX1Q, VTLQPX1G, VTLQPX14, VTLQPX0T, VTLQPX1Y, VTLQPX1M, VTLQPX2G, VTLQPX1O, VTLQPX1K, VTLQPX19, VTLQPX2E, VTLQPX1W, VTLQPX10, VTLQPX1Z, VTLQPX24, VTLQPX21, VTLQPX0X, VTLQPX1E, VTLQPX0U, VTLQPX26, VTLQPX2F, VTLQPX25, VTLQPX0Z, VTLQPX18, VTLQPX28, VTLQPX20, VTLQPX27, VTLQPX11, VTLQPX2B, VTLQPX1F, VTLQPX2H, VTLQPX41, VTLQPX3T, VTLQPX43, VTLRPX0R, VTKYPX09, VTKYPX44, VTKYPX46, VTKYPX47, VTKYPX48, VTKYPX49, VTKZPX14, VTKZPX15, VTKZPX20, VTKZPX17, VTLNPX4I, VTLNPX49, VTLNPX4G, VTLNPX4N, VTLNPX4E, VTLNPX4L, VTLNPX44, VTLNPX46, VTLNPX4C, VTLPPX27, VTLPPX28, VTLPPX29, VTLPPX1Z, VTLPPX37, VTLPPX3B, VTLPPX34, VTLPPX38, VTLPPX36, VTLQPX0Q, VTLQPX0B, VTLQPX0E, VTLQPX0P, VTLQPX0F, VTLRPX0X, VTLRPX0Z, VTKYPX07, VTKYPX06, VTKYPX04, VTLNPX4B, VTLNPX42, VTLNPX4F, VTLPPX3D, VTLQPX0M, VTKYPX51

View the original FDA recall notice

Why was this product recalled?

Breast biopsy and tissue removal system may experience the following errors: 1) Probe calibration and /or handheld driver failure (code HH01-00020009) , 2) Pressure sensor (code E01-0054000E), which prompts the user to restart and re-calibrate; errors may lead to interruption or prolongation of the biopsy procedure, or the need for additional tissue sampling or repeat biopsy.

What actions did Bard Peripheral Vascular, Inc. take?

On 8/20/2026, correction notices were mailed to customer who were asked to do the following: 1.If calibration cannot be completed, clinicians should first follow the troubleshooting instructions provided on the screen and in the Instructions for Use (IFU). If the issue persists, the probe should be replaced and the process repeated. If unsuccessful, an alternative biopsy driver or system should be used, as clinically appropriate. 2. If a pressure sensor error is observed, clinicians should follow the troubleshooting instructions provided on the screen and in the Instructions for Use (IFU). If the error occurs and the system cannot be returned to normal operations, discontinue use of the affected system and complete or reschedule using alternative equipment, as clinically appropriate. No additional follow-up is recommended for patients whose procedures were completed without complication. 3. If affected devices function as intended or if the error codes can be cleared, facilities can continue to use them prior to software update. 4. Post the correction notice on or near affected devices until the software has been updated. 5. Ensure devices are available for the update. 6. This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Share this notification with all product users within your facility network to ensure they are also aware of this product correction. If the affected products have been distributed further, please ensure that these entities are contacted and informed of the correction. 7. Complete and return the response form via email to BDRC33@bd.com Firm has developed a software solution to address these issues. A firm representative will contact your facility to arrange the update. If you require assistance, contact the firm via email: productcomplaints@bd.com, or phone: 1-844-823-5433

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology