Olympus M3 Gold Autoclavable Lateral Telescope 70. Recall
Olympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Olympus M3 Gold Autoclavable Lateral Telescope 70.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3300-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 24, 2026
- FDA report date
- Oct 7, 2026
- Quantity in commerce
- 50 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Olympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.
Affected model & UDI checker
Search the exact identifiers published for Z-3300-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| M3-70A | Olympus M3 Gold Autoclavable Lateral Telescope 70. | Check source notice | Not specified |
| Not specified | Olympus M3 Gold Autoclavable Lateral Telescope 70. | Check source notice | 00821925005686 |
FDA code information
Model Number: M3-70A. UDI-DI: 00821925005686. All Serial Numbers.
Why was this product recalled?
There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.
What actions did Olympus Corporation of the Americas take?
Olympus notified the single Japanese distributor on about 08/24/2026 via emailed "URGENT: FIELD SAFETY NOTICE." Consignees and end users are being instructed to store a copy of the provided correct sterilization parameters with current package insert, use only the corrected sterilization parameters, complete and return the provided reply form, and if units have been further distributed, notify those customers.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3300-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology