PALACOS R+G pro 40; PALACOS R+G pro 80. Recall
PALACOS R+G pro 40; PALACOS R+G pro 80. Model/Catalog Number: 5081288; 5081289. Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix).
FDA recall Class II
Ongoing
FDA lists this Class II recall for PALACOS R+G pro 40; PALACOS R+G pro 80.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3302-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 20, 2026
- FDA report date
- Oct 7, 2026
- Quantity in commerce
- 59,660 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
PALACOS R+G pro 40; PALACOS R+G pro 80. Model/Catalog Number: 5081288; 5081289. Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix).
Affected model & UDI checker
Search the exact identifiers published for Z-3302-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 5081288 | PALACOS R+G pro 40; PALACOS R+G pro 80. | Check source notice | Not specified |
| Not specified | PALACOS R+G pro 40; PALACOS R+G pro 80. | Check source notice | 04260102136182 |
FDA code information
Lot Code: UDI-DI: 04260102136182; 04260102136212. All Batch Numbers. All Lot Numbers.
Why was this product recalled?
Potential for monomer liquid leakage.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Heraeus Medical GmbH (Dental Division) take?
Heraeus Medical notified consignees on about 08/20/2026 via emailed letter. Consignees were informed of the issue and risk and were instructed to read the Correction Letter carefully and ensure that all relevant hospital departments are made aware of its contents, ensure that suitable eye protection is worn whenever the system is being operated, and complete and return the provided reply form. If any leakage is observed, do not continue to use the product. Instead, discard it with additional precautions, such as wearing double gloves (e.g. nitrile rubber), in accordance with internal hospital procedures and local waste disposal guidelines. Additionally, consignees were instructed to keep a copy of the provided response form with your records in case your organization is subject to a compliance audit.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3302-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.