Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 Recall
Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 Model/Catalog Number: HX55W-S0 Software Version: N/A Product Description: The Quantum PureFlow Standard Heat Exchanger is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration with a maximum flow rate of 8 l/min. Quantum PureFlow Standard Heat Exchanger is a single use device which allows heating/cooling the blood during extracorporeal circulation. The device is sold sterile in a blister pack in a multiple box of 4 units. Device surfaces in contact with blood are treated with a phosphorylcholine-based biocompatible coating. The device has been designed to be powered by heaters coolers that use water or a glycol-based heat transfer fluid (HTF). Component: N/A
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3328-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jul 17, 2026
- FDA report date
- Oct 7, 2026
- Quantity in commerce
- 3232
- Distribution
- See key details
- FDA cause
- Process design
What is being recalled?
Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 Model/Catalog Number: HX55W-S0 Software Version: N/A Product Description: The Quantum PureFlow Standard Heat Exchanger is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration with a maximum flow rate of 8 l/min. Quantum PureFlow Standard Heat Exchanger is a single use device which allows heating/cooling the blood during extracorporeal circulation. The device is sold sterile in a blister pack in a multiple box of 4 units. Device surfaces in contact with blood are treated with a phosphorylcholine-based biocompatible coating. The device has been designed to be powered by heaters coolers that use water or a glycol-based heat transfer fluid (HTF). Component: N/A
Affected model & UDI checker
Search the exact identifiers published for Z-3328-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| HX55W-S0 | Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 | Check source notice | Not specified |
FDA code information
Lot Code: 18051160301178 (HX55W-S0)
Why was this product recalled?
Due to blood-to-glycol leak in the heat exchanger.
What actions did SPECTRUM MEDICAL SRL take?
On 07/17/2026, an "URGENT MEDICAL DEVICE RECALL" letter was sent via email to customers informing them that, Spectrum Medical S.r.l. has received customer complaints reporting a leak between the blood and glycol compartments of certain adult heat exchangers. Actions to be taken by customers: During the procedure, the user should monitor the heat exchanger for leakage, which would be evident by a change in colour in the glycol compartment. Below are images of the heat exchanger if a blood-to-glycol leak was to occur. When operating as intended, the glycol compartment should remain clear. If the colour of the glycol compartment changes, then a leak is present between the blood and glycol compartments. If leakage is noted between the blood and glycol compartments of the heat exchanger, the heat exchanger should be immediately replaced in the circuit with a new device. Please return the defective heat exchanger to Spectrum Medical. The contaminated heater-cooler device should also be immediately removed from clinical use and returned to Spectrum Medical for decontamination and inspection. Do not use the heater-cooler with another patient. If leakage has occurred, please contact a Spectrum Medical representative to coordinate device return. If leakage into the water compartment occurs when using a third-party heater-cooler, the customer should immediately remove the heater-cooler from clinical use and contact the heater-cooler manufacturer. Spectrum Medical suggests that for customers that use the heat exchanger in combination with a third-party heater-cooler to contact the manufacturer of the heater-cooler to pre-emptively discuss the steps that should be taken if the device is exposed to blood contamination. For questions or further assistance - contact Regulatory Affairs Specialist at cameron.schmitt@spectrummedical.com
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3328-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.