HOPKINS Telescope 30, 4mm REF 10005BA UDI-DI code: 04048551045121 KARL STORZ HOPKINS and HOPKINS II telescopes are to be used for visualization of the surgical site during diagnostic and therapeutic endoscopic procedures. Recall
HOPKINS Telescope 30¿, 4mm REF 10005BA UDI-DI code: 04048551045121 KARL STORZ HOPKINS¿ and HOPKINS¿ II telescopes are to be used for visualization of the surgical site during diagnostic and therapeutic endoscopic procedures.
FDA recall Class II
Ongoing
FDA lists this Class II recall for HOPKINS Telescope 30, 4mm REF 10005BA UDI-DI code: 04048551045121 KARL STORZ HOPKINS and HOPKINS II telescopes are to be used for visualization of the surgical site during diagnostic and therapeutic endoscopic procedures.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3333-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 12, 2026
- FDA report date
- Oct 7, 2026
- Quantity in commerce
- 6 eyepieces
- Distribution
- See key details
- FDA cause
- Employee error
What is being recalled?
HOPKINS Telescope 30, 4mm REF 10005BA UDI-DI code: 04048551045121 KARL STORZ HOPKINS and HOPKINS II telescopes are to be used for visualization of the surgical site during diagnostic and therapeutic endoscopic procedures.
Affected model & UDI checker
Search the exact identifiers published for Z-3333-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 10005BA | HOPKINS Telescope 30, 4mm REF 10005BA UDI-DI code: 04048551045121 KARL STORZ HOPKINS and HOPKINS II telescopes are to be used for visualization of the surgical site during diagnostic and therapeutic endoscopic procedures. | Check source notice | Not specified |
| Not specified | HOPKINS Telescope 30, 4mm REF 10005BA UDI-DI code: 04048551045121 KARL STORZ HOPKINS and HOPKINS II telescopes are to be used for visualization of the surgical site during diagnostic and therapeutic endoscopic procedures. | Check source notice | 04048551045121 |
FDA code information
Material Number: 10005BA UDI-DI code: 04048551045121 Serial Numbers: 100011, 100012, 10000Y, and 10000Z
Why was this product recalled?
Due to incorrect inside packaging label for eyepiece.
What actions did Karl Storz Endoscopy take?
On 08/12/2026, an "URGENT: MEDICAL DEVICE RECALL" Letter was sent via email informing customers that, KARL STORZ distributed 6 HOPKINS Telescope 30, 4mm REF: 10005BA between November 24, 2025, and December 15, 2025, that contains an optic (eyepiece) that is incorrectly etched with REF: 10005AA, which is a HOPKINS Telescope 0, 4 mm. The optic (eyepiece) inside the packaging is in fact a REF: 10005BA HOPKINS Telescope 30, 4mm, as the outer packaging indicates. Customers are instructed to: 1. Immediately check for impacted serial numbers in inventory and quarantine the incorrectly etched devices for return to Karl Storz. 2. Pass on this Urgent Medical Device Correction to all users of the products listed below (Table 1 and Table 2) and all other persons who need to be aware within your organization. 3. If customers have or may have distributed the products listed, identify and promptly notify those recipients with a copy of this notice and response form. 4. Return the completed Customer Reply Form by E-Mail to the indicated contact below within 15 calendar days from the date of receipt. 5. Report any incidents related to this issue to the manufacturer, dealer, or local representative and, if applicable, to the national competent authority, as this is important feedback. For questions - email to KSEAProductLabeling@karlstorz.com.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA classification recorded
FDA notice posted
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3333-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology