FDA enforcement records identify Beckman Coulter Mishima K.K. as the recalling firm for recall Z-3332-2026, classified Class II, concerning the BECKMAN COULTER CHEMISTRY ANALYZER AU480. The agency's device recall database lists the action as a Class 2 device recall, and the enforcement entry records the firm's location as Sunto-Gun, Japan, with the recall initiated on August 3, 2026 and reported on October 7, 2026. The manufacturer's own product page describes the AU480 as a compact clinical chemistry analyzer, and the recall record states the system measures analytes in samples for in vitro diagnostic use in combination with reagents, calibrators, quality control material and other accessories. [1][2][6]

What the recall covers

The recall record describes products containing affected software, identified by item numbers B96692, B96693 and B12183, each with a listed UDI-DI code, plus item number N3151400, for which the record states no UDI number is available because the device exceeded its storage period. The stated product quantity is 492 units. The reason given is that certain AU480 analyzers with Windows 10 IoT Enterprise 2021 LTSC PCs may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software. [1]

A secondary summary of the FDA notice states that when this occurs, AlarmDB becomes unreadable and the instrument software cannot be started, and that the affected configurations involve Windows 10 IoT Enterprise 2021 LTSC PCs pre-installed with Microsoft SQL Server 2022 Express. The same summary characterizes the action as moderate risk. [3][7]

Status and distribution

The enforcement entry lists the recall status as Ongoing, with initial firm notification by letter. The FDA device recall page notes that not all products have been corrected or removed and that the record will be updated as the status changes; that page carries a last-updated date of October 30, 2026. [1][2]

The distribution field in the enforcement record names U.S. states including CA, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, MN, MO, NC, NM, NY, OK, PA, SD, TX and WI, and countries including Algeria, Australia, Belarus, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Cyprus, Czechia, Egypt, France, Germany, Guatemala, Hungary, India, Ireland, Italy, Japan, Kazakhstan, Korea, Republic of, Latvia, Malaysia, Morocco, Myanmar, Nepal, Nigeria, Pakistan, Poland, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, South Africa, Taiwan Province of China, Turkey, United Arab Emirates and Viet Nam. [1]

The AU480 system as documented

The manufacturer's published product page presents the AU480 Chemistry Analyzer as a compact, reliable and cost-effective clinical chemistry analyzer, and states that instructions for use must be consulted before using these products. The recall record's product description states that the AU480 measures analytes in samples, in combination with appropriate reagents, calibrators, quality control material and other accessories, and is for in vitro diagnostic use only. [6][1]

FDA establishment registration and device listing records show Beckman Coulter Mishima K.K. under registration number 9612296 for product code JJE, with that listing page last updated October 12, 2026. The recall record identifies the recalling firm by the same name. [9][1]

How to read the record

The recall number Z-3332-2026 links the enforcement entry, the FDA device recall page and the event record under event ID 99588. Third-party recall trackers reproduce the FDA notice and describe themselves as independent and not affiliated with any government agency; their summaries should be checked against the agency record. [1][2][4][7]

The record documents a software-related failure mode and a distribution scope; it does not state clinical performance, corrective action completion or a root cause beyond the described software configuration. Those points remain unknown from the cited documents. [1][2]