Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-0164-2027

TS Universal Baseplate Finishing Tray, Triathlon. Recall

TS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506.

FDA recall Class II

Ongoing

FDA lists this Class II recall for TS Universal Baseplate Finishing Tray, Triathlon.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-0164-2027
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Sep 16, 2026
FDA report date
Oct 21, 2026
Quantity in commerce
605 trays
Distribution
See key details
FDA cause
Labeling design

What is being recalled?

TS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506.

Affected model & UDI checker

Search the exact identifiers published for Z-0164-2027.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
6543-6-506.TS Universal Baseplate Finishing Tray, Triathlon.Check source notice07613327548082
6543-6-506TS Universal Baseplate Finishing Tray, Triathlon.Check source noticeNot specified
FDA code information

Catalog Number: 6543-6-506. GTIN: 07613327548082. All lot numbers used prior to the correction date of May 8, 2026 are impacted.

View the original FDA recall notice

Why was this product recalled?

Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.

What actions did HOWMEDICA OSTEONICS CORP take?

Stryker notified consignees on about 09/16/2026 via "MEDICAL DEVICE CORRECTION" letter mailed through UPS 2-day air. Consignees were informed that reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes, however, these trays require a sterilization dry time of 40 minutes when sterilized in blue wrap. Consignees were instructed to inform users of the Medical Device Correction and forward the notice to all individuals who need to be made aware or organizations who have consigned product, review and follow the updated instructions provided in LSTPI-B Cleaning, Sterilization, Inspection, and Maintenance of Reusable Medical Devices Reference Guide, complete and return the provided Medical Device Correction Business Reply Form.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology