FDA device enforcement records identify HOWMEDICA OSTEONICS CORP as the recalling firm for two distinct correction events. Event 99883 covers four Triathlon instrument trays classified Class II, with a stated reason tied to reprocessing instructions in LSTPI-B specifying a 30-minute sterilization dry time for trays sterilized in blue wrap. Event 99558 covers Scorpio, Mobile Bearing PCS and System One Bipolar orthopaedic implant components made from UHMWPE raw material aged beyond five years. The FDA establishment listing for the Mahwah, New Jersey address records the firm as active and as an initial distributor or importer, with trade names Stryker and Stryker Orthopaedics. [1][5][9]

Triathlon tray correction and its stated cause

Four Triathlon trays appear in the enforcement data under recall numbers Z-0164-2027 through Z-0167-2027: the TS Universal Baseplate Finishing Tray (6543-6-506), TS Femoral Prep Tray (6543-6-507), Triathlon Hinge Femoral Prep Tray (6543-6-514) and TS Misc. Instruments Tray (6543-6-519). Each is Class II, firm-initiated and voluntary, with an initiation date of September 16, 2026 and a center classification date of October 15, 2026. The stated reason is that the 30-minute dry time in LSTPI-B may leave a small amount of residual moisture after sterilization when blue wrap is used. [1][2][3][4]

Stated quantities differ by tray: 605 for the baseplate finishing tray, 582 for the femoral prep tray, 359 for the hinge femoral prep tray and 577 for the miscellaneous instruments tray. Each record states that all lot numbers used prior to the May 8, 2026 correction date are impacted, that trays are provided by sale or loan, and that 158 trays were distributed outside the United States. The FDA recall database page for event 99883 returns four related results. [1][2][3][4][10]

UHMWPE implant components and stated hazards

Recall Z-3205-2026 covers Scorpio and Mobile Bearing PCS components plus System One Bipolar heads in 40 mm, 43 mm and 44 mm sizes, identified by catalog numbers 78-5-0910, 6074-4022, 6074-4322 and 6074-4422. The record states that affected devices were manufactured with UHMWPE raw material aged in excess of five years, with the potential for elevated oxidation and increased wear. Listed potential hazards include excessive wear debris, material fragments, fractured or disassociated devices, revision surgery, pain and inflammation. [5]

The filing is Class II and voluntary, initiated July 30, 2026 and classified September 16, 2026, with a stated quantity of 12,156 units in the United States. Distribution is described as worldwide, covering the United States and the countries of Canada, LATAM, Europe, Asia Pacific and South Africa. Lot and UDI-DI identifiers are listed for each catalog number in the record. [5]

Applicant identity and published registration data

The FDA establishment registration and device listing entry names HOWMEDICA OSTEONICS CORP at 325 Corporate Dr., Mahwah, NJ 07430, with registration number 2249697 and FEI number 3003070421. The entry lists business trade names Stryker and Stryker Orthopaedics, records status as active, and marks the firm as an initial distributor or importer. That registration address is a filing address and is not presented here as a verified headquarters. [9]

The enforcement records for both events list the same Mahwah address and name Howmedica Osteonics Corp. as the recalling firm. The Triathlon tray records carry GTIN identifiers for each catalog number, and the UHMWPE record carries UDI-DI identifiers. The FDA recall pages for the Triathlon hinge and femoral prep trays confirm Class 2 designation in their titles. [1][2][5][6][8]

How to read the filing record

The two events are separate: one concerns instrument tray reprocessing instructions, the other concerns implant raw-material aging. Both are recorded as ongoing at the time of the enforcement entries. The Triathlon tray records share event 99883 and the same stated reason, while the UHMWPE record carries event 99558 and a different stated cause. [1][5]

No submission clearances, sales figures, market totals or safety performance measures for this firm appear in the numbered records. The quantities above are those stated in individual enforcement entries, not company-wide volumes. [1][5]