The public record for this firm is narrow. FDA enforcement record Z-0356-2027, event 99959, lists CORIN MEDICAL, LTD. as the recalling firm for the Unity Knee ASC Modular Handle, part number 1232040, with lot code 097921 and UDI-DI/GTIN 05056139239656. The record describes the product as part of the Corin UNITY ASC INSTRUMENT SYSTEM and states a potential malfunction in which the handle’s locking mechanism may fail, leaving engaged mating instruments unable to disengage and preventing secure fitting to unengaged instruments. The recall is classified Class II, marked voluntary and firm initiated, and listed as ongoing. The record gives a Cirencester, United Kingdom address for the firm, which is a filing address and not evidence of a headquarters. The numbered sources do not establish a corporate relationship between this recalling firm and CORIN LIMITED, the manufacturer named in the Unity Total Knee System instructions for use. [1][3][4]
The recalled instrument and its stated failure mode
The FDA enforcement entry identifies the affected article as the Unity Knee ASC Modular Handle, part number 1232040, within the Corin UNITY ASC INSTRUMENT SYSTEM. The stated reason for recall is a potential malfunction of the modular handles in which the locking mechanism may fail; mating instruments already engaged to the handle would not be able to be disengaged, and the handle would be unable to fit securely to unengaged mating instruments. The record lists 22 units, lot code 097921, and worldwide distribution including US nationwide distribution in the states of CA, GA, IL, MI, NC, RI, TX and UT. [1]
The FDA recall database entry for the same action records the classification as Class 2 and states that Corin notified consignees on about 10/02/2026 through an emailed “URGENT MEDICAL DEVICE REMOVAL” letter. It also records the cause as FDA determined, under investigation by the firm. The enforcement record gives a report date of 20261111 and a status of ongoing. [3][1]
Filing identifiers and dates
The enforcement record carries recall number Z-0356-2027 and event number 99959, with the firm identified as CORIN MEDICAL, LTD. and the product type listed as devices. The recall database entry gives the firm FEI as 3002806725 and the product code as JWH, with the action initiated 2026-10-02 and posted 2026-11-03. These identifiers describe the enforcement action, not a marketing authorization for an implant. [1][5]
The same database entry lists a series of 510(k) numbers associated with the record, spanning K790219 through K202429. The numbered sources do not state which of those clearances cover the modular handle or the wider instrument system, so no clearance is attributed here to a specific product. [5]
Published Unity Knee documentation
The coringroup.com Unity Knee page links operative technique documents for the Unity Knee with MC, an ASC instrumentation operative technique, a difficult primary indications technique, an evidence base document, a product overview, and product flyers, including one described as streamlined instrumentation for total knee replacement. Those links are published by the site’s operator; the numbered sources do not show that the operator is the same legal entity as the recalling firm. [2]
A Unity Total Knee System instructions for use names CORIN LIMITED, The Corinium Centre, Cirencester, Gloucestershire, as manufacturer and states that the document has been approved by the U.S. Food and Drug Administration. It directs that implantation be performed only by a qualified surgeon familiar with the specific operative technique, and references ASTM F2503-20 for marking in the magnetic resonance environment. This document is a manufacturer publication and does not name CORIN MEDICAL, LTD. [4]
What the record does not establish
The numbered sources do not document total sales, total portfolio size, safety performance across products, customers, certifications, or commercial availability of any Corin product. The single enforcement record and the linked product documentation describe one instrument component and one knee system’s published materials; they do not support broader conclusions about the firm’s device range or quality history. [1][2][4]