Medical devices, healthcare technology and the business of care
COMPANY INTELLIGENCE

Medtronic Neuromodulation

Explore this company’s associated devices, FDA decisions and regulatory history.

Clinical monitoring equipment, editorial photography

Medtronic Neuromodulation

Regulatory records
130
Linked recall entries
14
Headquarters
Not reported
Brand website
medtronic.com
COMPANY INTELLIGENCE

What the linked records show

MedtechLedger currently links 130 FDA records to the applicant name Medtronic Neuromodulation. It includes 4 original PMA approvals. 28 entries are PMA supplements. 98 entries are 30-day notices. These are records held by this publication, rather than a count of the company’s entire portfolio.

The dated entries run from Nov 30, 1984 to Oct 30, 2026. Individual filings may concern different products or changes to an existing application; review the identifier and decision scope on each record.

Product code LGW appears in 41 of the 130 linked filings. The collection spans 4 FDA product codes and 4 original PMA application families. These figures describe the filings available here; they do not establish the company’s total device sales or market share.

The largest PMA family in this collection is P840001, with 41 linked entries. Follow the parent application and its dated supplements to distinguish the original device scope from later changes. 16 linked filings have a current plain-language explanation checked against the recorded FDA decision.

Applicant mailing address recorded by FDA: 7000 Central Ave., NE, Minneapolis, MN, 55432. Recorded on P840001/S636, decision dated Oct 30, 2026.

Changes described in PMA records

P840001/S636: Restore, Itrel, Synergy, Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces-Quad, Specify,

FDA recorded decision OK30 on October 30, 2026 for Medtronic Neuromodulation's P840001/S636. This 30-Day Notice concerns production changes: an additional surface mount technology (SMT) line, a revised manufacturing sequence, and inspection and de-paneling equipment for printed circuit board assembly production. The database places the entry under process changes involving a manufacturer, sterilizer, packager or supplier.

The statement specifies that automated optical inspection will occur immediately after each SMT assembly stage. It also describes an equivalent solder paste inspection system and a qualified de-paneling machine. These are the particular factory steps and equipment identified in this notice. The record does not supply machine specifications, inspection acceptance limits, validation results or the full production procedure. It identifies product code LGW and the Neurology specialty, but provides no separate clinical findings for these manufacturing changes.

FDA approval-order wording

approval for the following three changes: 1) Adding a new surface mount technology (SMT) production line.2) Revision of the SMT manufacturing process sequence to perform automated optical inspection immediately following each SMT assembly stage.3) Addition of an equivalent solder paste inspection system and addition of a qualified de-paneling machine for PCBA production

Oct 30, 2026 · 30-Day Notice

P970004/S460: InterStim Therapy System, Verify Evaluation System (SNS Urinary Neurostimulator 97800), (SNS Urinary Neurostimulator I

The FDA recorded decision OK30 on 2026-10-29 for PMA supplement P970004/S460, submitted by Medtronic Neuromodulation as a 30-Day Notice for a process change affecting manufacturing, sterilization, packaging or supplier operations. The supplement sits under P970004, whose listed title covers the InterStim Therapy System and Verify Evaluation System, including the SNS Urinary Neurostimulator 97800. The device category is an electrical implantable stimulator for incontinence, with gastroenterology and urology specialties and product code EZW.

The recorded statement describes battery-line process controls rather than a new approval of the whole system: enhanced operator verification of the FactoryWorks step and recorded battery pre-weight, pre- and post-weight out-of-control action plans, software and human machine interface updates that reject batteries above the established final-fill reasonable limit, dedicated white work-in-progress trays, an isopropyl alcohol wipe for residual electrolyte, and removal of Ball Press and Button Press certification prerequisites for ram changes on Medium Rate Line 2, with supervisors and engineers retaining certification responsibility. No documents were provided, so the underlying supplement filing and any clinical or labeling effect remain unknown.

The listed date is in the future relative to typical publication timelines, and the source does not state whether the notice was accepted without further review or what follow-up applies. Whether these manufacturing changes alter device performance or labeling cannot be determined from the record.

FDA approval-order wording

1. Enhanced operator verification of the applicable FactoryWorks processing step and recorded battery pre-weight.2. Addition of pre-weight and post-weight out-of-control action plans to the applicable Process Operation Descriptions.3. Software and Human Machine Interface updates that will automatically reject batteries with fill weights above the established final-fill reasonable limit.4. Introduction of dedicated white work-in-progress trays to distinguish electrolyte-filled batteries and improve line clearance, handling, and traceability.5. Use of an isopropyl alcohol wipe to remove residual electrolyte from battery surfaces before subsequent processing.6. Removal of the Ball Press and Button Press certification prerequisites from the Ball and Button Press Ram Change procedure, allowing Medium Rate Line 2 operators to qualify for press ram changes by completing only the ram-change procedure while supervisors and engineers retain responsibility for operator certification

Oct 29, 2026 · 30-Day Notice

P080025/S354: InterStim Therapy System, Verify Evaluation System (SNS Bowel Neurostimulator 97800), (SNS Bowel Neurostimulators Impl

FDA posted a 30-day notice for supplement S354 to PMA P080025, submitted by Medtronic Neuromodulation for the InterStim Therapy System and Verify Evaluation System. The supplement reason is a process change involving the manufacturer, sterilizer, packager, or supplier. The decision field records OK30, and the product code is QON for an implanted electrical device intended for treatment of fecal incontinence.

The statement lists six manufacturing changes: enhanced operator verification of a FactoryWorks processing step and recorded battery pre-weight; addition of pre-weight and post-weight out-of-control action plans to Process Operation Descriptions; software and human machine interface updates that automatically reject batteries above the final-fill reasonable limit; dedicated white work-in-progress trays for electrolyte-filled batteries; an isopropyl alcohol wipe to remove residual electrolyte; and removal of Ball Press and Button Press certification prerequisites from the ram-change procedure, with supervisors and engineers retaining responsibility for operator certification. The notice does not establish safety or clinical effectiveness.

FDA approval-order wording

1. Enhanced operator verification of the applicable FactoryWorks processing step and recorded battery pre-weight.2. Addition of pre-weight and post-weight out-of-control action plans to the applicable Process Operation Descriptions.3. Software and Human Machine Interface updates that will automatically reject batteries with fill weights above the established final-fill reasonable limit.4. Introduction of dedicated white work-in-progress trays to distinguish electrolyte-filled batteries and improve line clearance, handling, and traceability.5. Use of an isopropyl alcohol wipe to remove residual electrolyte from battery surfaces before subsequent processing.6. Removal of the Ball Press and Button Press certification prerequisites from the Ball and Button Press Ram Change procedure, allowing Medium Rate Line 2 operators to qualify for press ram changes by completing only the ram-change procedure while supervisors and engineers retain responsibility for operator certification

Oct 29, 2026 · 30-Day Notice

P840001/S635: Master Restore, Itrel, Synergy Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces, Specify,

FDA posted a 30-day notice for PMA supplement S635 to P840001, covering Medtronic Neuromodulation spinal cord stimulation systems and leads. The dated statement lists six manufacturing process changes: enhanced operator verification of a FactoryWorks processing step and recorded battery pre-weight; addition of pre-weight and post-weight out-of-control action plans to Process Operation Descriptions; software and human machine interface updates that automatically reject batteries with fill weights above the established final-fill reasonable limit; introduction of dedicated white work-in-progress trays to distinguish electrolyte-filled batteries; use of an isopropyl alcohol wipe to remove residual electrolyte from battery surfaces; and removal of Ball Press and Button Press certification prerequisites from the ram change procedure.

The notice is a process change supplement, not a new approval of the whole system or an original PMA. The statement reports the intended purpose of the changes, such as improving line clearance, handling, and traceability, but does not independently establish safety or clinical effectiveness. No clinical indication, trial data, or benefit assessment is included. The source does not state whether these changes affect device labeling or performance specifications.

FDA approval-order wording

1. Enhanced operator verification of the applicable FactoryWorks processing step and recorded battery pre-weight.2. Addition of pre-weight and post-weight out-of-control action plans to the applicable Process Operation Descriptions.3. Software and Human Machine Interface updates that will automatically reject batteries with fill weights above the established final-fill reasonable limit.4. Introduction of dedicated white work-in-progress trays to distinguish electrolyte-filled batteries and improve line clearance, handling, and traceability.5. Use of an isopropyl alcohol wipe to remove residual electrolyte from battery surfaces before subsequent processing.6. Removal of the Ball Press and Button Press certification prerequisites from the Ball and Button Press Ram Change procedure, allowing Medium Rate Line 2 operators to qualify for press ram changes by completing only the ram-change procedure while supervisors and engineers retain responsibility for operator certification

Oct 29, 2026 · 30-Day Notice

P970004/S459: InterStim Therapy System, Verify Evaluation System

The FDA record P970004/S459, decided 2026-10-09 as OK30 under a 30-Day Notice for a process change involving a manufacturer, sterilizer, packager, or supplier, covers the InterStim Therapy System and Verify Evaluation System. The stated purpose is approval to introduce a second Medium Rate (MR) Header manufacturing line, MR Line 2, and to make inspection updates to the MR Headers Line currently in production at Medtronic Energy and Components Center (MECC), Brooklyn Center, MN. The applicant is Medtronic Neuromodulation; the device category is an electrical implantable stimulator for incontinence, product code EZW.

This is a supplement to PMA P970004, not a new approval of the whole system, and the notice concerns manufacturing and inspection at one site. The record does not describe header specifications, the inspection changes, validation data, or any clinical or safety finding; the stated purpose alone does not establish safety or effectiveness.

FDA approval-order wording

approval for the introduction of a second Medium Rate (MR) Headers manufacturing line (MR Line 2) and to make inspection updates to the MR Headers Line that is currently in production at Medtronic Energy and Components Center (MECC), Brooklyn Center, MN.

Oct 9, 2026 · 30-Day Notice

Subsidiaries and similarly named applicants can have separate records. The links below preserve the applicant and submission identifiers. Source records are linked on each filing.

RECALL COVERAGE

Recalls linked to this firm

MedtechLedger links 14 FDA enforcement entries through the recorded recalling firm or an exact submission number published in FDA’s recall data. 14 Class II records are represented. Each row is an enforcement entry; one recall event can cover several product entries. These counts describe this collection and do not measure a manufacturer’s safety record or an injury rate.

Recall / productCompanyClassificationReport dateSource
Medtronic NeuromodulationClass IIOct 28, 2026FDA
Medtronic NeuromodulationClass IIOct 28, 2026FDA
Medtronic NeuromodulationClass IIJun 3, 2026FDA
Medtronic NeuromodulationClass IIJun 3, 2026FDA
Medtronic NeuromodulationClass IIMay 20, 2026FDA
Medtronic NeuromodulationClass IIJan 21, 2026FDA

Connections use the recorded recalling firm or the submission identifiers shown below. A shared submission connects the records; it does not establish that differently named legal applicants are interchangeable. Use the exact affected models, lots and instructions on each notice.

Search the firm’s recall records
Source information

Source updated Nov 11, 2026. Retrieved Nov 21, 2026.

Regulatory record history

  1. Restore, Itrel, Synergy, Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces-Quad, Specify,

    P840001/S636 · PMA supplement
  2. InterStim Therapy System, Verify Evaluation System (SNS Urinary Neurostimulator 97800), (SNS Urinary Neurostimulator I

    P970004/S460 · PMA supplement
  3. InterStim Therapy System, Verify Evaluation System (SNS Bowel Neurostimulator 97800), (SNS Bowel Neurostimulators Impl

    P080025/S354 · PMA supplement
  4. Master Restore, Itrel, Synergy Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces, Specify,

    P840001/S635 · PMA supplement
  5. InterStim Therapy System, Verify Evaluation System

    P970004/S459 · PMA supplement
  6. InterStim Therapy System, Verify Evaluation System

    P080025/S353 · PMA supplement
  7. lnterStim Therapy System, Verify Evaluation System

    P080025/S351 · PMA supplement
  8. InterStim Therapy System, Verify Evaluation System

    P970004/S457 · PMA supplement
  9. Master Vanta with Restore, Itrel, Synergy, Intellis, AdaptiveStim Vanta and Inceptiv TM Spinal Cord Implantable S

    P840001/S632 · PMA supplement
  10. Vantawith AdaptiveStim Technology - Model 977006

    P840001/S634 · PMA supplement
  11. InterStim Therapy System, Verify Evaluation System

    P080025/S352 · PMA supplement
  12. InterStim Therapy System Verify Evaluation System

    P970004/S458 · PMA supplement
  13. Altaviva

    P240011/S016 · PMA supplement
  14. Inceptiv Implantable Neurostimulator, Vanta with AdaptiveStim Implantable Neurostimulator, Inceptiv' LT Implantable N

    P840001/S633 · PMA supplement
  15. InterStim Therapy System, Verify Evaluation System

    P970004/S456 · PMA supplement
  16. InterStim Therapy System, Verify Evaluation System

    P080025/S350 · PMA supplement
  17. Altaviva

    P240011/S015 · PMA supplement
  18. Master Restore, Itrel, Synergy, Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces-Quad, Spe

    P840001/S631 · PMA supplement
  19. InterStim Therapy System, Verify Evaluation System

    P970004/S455 · PMA supplement
  20. Master Restore; Itrel; Synergy; Intellis; Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces; Specify

    P840001/S630 · PMA supplement
  21. Altaviva

    P240011/S010 · PMA supplement
  22. Master Restore, Itrel, Synergy, Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces, Specify

    P840001/S623 · PMA supplement
  23. InterStim Therapy System, Verify Evaluation System

    P970004/S454 · PMA supplement
  24. Altaviva

    P240011/S014 · PMA supplement
  25. InterStim Therapy System, Verify Evaluation System

    P080025/S349 · PMA supplement
  26. Master Restore, Itrel, Synergy, Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces-Quad, Spe

    P840001/S629 · PMA supplement
  27. InterStim Therapy System

    P080025/S348 · PMA supplement
  28. InterStim Therapy System

    P970004/S453 · PMA supplement
  29. Master Restore; Itrel; Synergy; Intellis; Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces; Specify

    P840001/S628 · PMA supplement
  30. InterStim TM Therapy System; Verify TM Evaluation System (97800)

    P080025/S347 · PMA supplement
  31. InterStim TM Therapy System; Verify TM Evaluation System (97800)

    P970004/S452 · PMA supplement
  32. InterStim TM Therapy System; Verify Evaluation System; InterStim SNS Bowel Lead Model 978B1

    P080025/S346 · PMA supplement
  33. InterStim Therapy System; Verify Evaluation System; InterStim SNS Urinary Lead Model 978B1

    P970004/S451 · PMA supplement
  34. InterStim Therapy System; Verify Evaluation System (3550-05, 3560031, 3889, 978A1, 978B1)

    P970004/S450 · PMA supplement
  35. Master Restore, Itrel, Synergy Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces, Specify,

    P840001/S626 · PMA supplement
  36. Master Restore; Itrel; Synergy; Intellis; Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces; Specify

    P840001/S627 · PMA supplement
  37. InterStim Therapy System; Verify Evaluation System (3550-05, 3560031, 3889, 978A1, 978B1)

    P080025/S345 · PMA supplement
  38. Master Restore, Itrel, Synergy, Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces, Specify

    P840001/S620 · PMA supplement
  39. Altaviva

    P240011/S013 · PMA supplement
  40. Master Restore, Itrel, Synergy Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces, Specify,

    P840001/S625 · PMA supplement
  41. InterStim Therapy System, Verify Evaluation System

    P970004/S448 · PMA supplement
  42. Master Restore, Itrel , Synergy, Intellis , Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces-Quad, Specify,

    P840001/S624 · PMA supplement
  43. InterStim Therapy System, Verify Evaluation System

    P080025/S343 · PMA supplement
  44. InterStim Therapy System, Verify Evaluation System

    P970004/S449 · PMA supplement
  45. InterStim Therapy System, Verify Evaluation System

    P080025/S344 · PMA supplement
  46. Altaviva

    P240011/S012 · PMA supplement
  47. InterStim Therapy System, Verify Evaluation System (SNS Urinary Implantable Neurostimulators 97800, 97810)

    P970004/S447 · PMA supplement
  48. Master Intellis , Vanta and Inceptiv Spinal Cord Stimulation Systems (SCS Implantable Neurostimulators 97716, 97715, 9

    P840001/S621 · PMA supplement
  49. InterStim Therapy System, Verify Evaluation System (SNS Bowel Implantable Neurostimulators 97800, 97810)

    P080025/S342 · PMA supplement
  50. Altaviva

    P240011/S011 · PMA supplement
  51. InterStim Therapy System, Verify Evaluation System (SNS Urinary Neurostimulator)

    P970004/S446 · PMA supplement
  52. InterStim Therapy System, Verify Evaluation System (SNS Bowel Neurostimulator)

    P080025/S341 · PMA supplement
  53. Altaviva

    P240011/S008 · PMA supplement
  54. InterStim Therapy System, Verify Evaluation System

    P970004/S445 · PMA supplement
  55. InterStim Therapy System, Verify Evaluation System

    P080025/S340 · PMA supplement
  56. Master Restore, Itrel, Synergy Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces, Specify,

    P840001/S619 · PMA supplement
  57. InterStim Therapy System, Verify Evaluation System (SNS Urinary Programming Systems A510, A511, A520, A521)

    P970004/S440 · PMA supplement
  58. InterStim Therapy System, Verify Evaluation System (SNS Bowel Programming Systems A510, A511, A520, A521)

    P080025/S335 · PMA supplement
  59. Master Restore, Itrel, Synerg, Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces, Specify, and Vec

    P840001/S615 · PMA supplement
  60. Restore, Itrel, Synergy, Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces, Specify, and V

    P840001/S609 · PMA supplement
  61. SNS Urinary Neurostimulator Models 97800, 97810; SNS Urinary Screening Trialing Systems Model 3531

    P970004/S444 · PMA supplement
  62. SNS Bowel Neurostimulator Models 97800, 97810; SNS Bowel Screening Trialing System Model 3531

    P080025/S339 · PMA supplement
  63. Altaviva Model P7850N

    P240011/S006 · PMA supplement
  64. External Neurostimulator (ENS) Model 97725; Intellis Model 97716; Intellis Model 97715; SCS Implantable Neurostimulator

    P840001/S617 · PMA supplement
  65. InterStim Therapy System, Verify Evaluation System

    P080025/S332 · PMA supplement
  66. Restore, Itrel, Synergy, IntellisTM and VantaTM Spinal Cord Stimulation Systems and PiscesTM, Specify, and Vectris S

    P840001/S608 · PMA supplement
  67. InterStim Therapy System, Verify Evaluation System

    P970004/S437 · PMA supplement
  68. Master Restore, Itrel, Synergy Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces, Specify,

    P840001/S618 · PMA supplement
  69. Altaviva Tibial Neuromodulation System (Altaviva P7850N)

    P240011/S007 · PMA supplement
  70. InterStim TM Therapy System, Verify Evaluation System

    P080025/S319 · PMA supplement
  71. InterStim TM Therapy System, Verify Evaluation System

    P970004/S424 · PMA supplement
  72. Altaviva

    P240011/S005 · PMA supplement
  73. InterStim Therapy System, Verify Evaluation System

    P970004/S443 · PMA supplement
  74. InterStim Therapy System, Verify Evaluation System

    P080025/S338 · PMA supplement
  75. InterStim Therapy System, Verify Evaluation System

    P080025/S337 · PMA supplement
  76. InterStim Therapy System, Verify Evaluation System

    P970004/S442 · PMA supplement
  77. InterStim Therapy System; Verify Evaluation System

    P970004/S441 · PMA supplement
  78. InterStim Therapy System; Verify Evaluation System

    P080025/S336 · PMA supplement
  79. Master Restore; Itrel; Synergy; Intellis; Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces; Specify; and Ve

    P840001/S616 · PMA supplement
  80. Medtronic Altaviva System

    P240011/S004 · PMA supplement
  81. Master Restore, Itrel, Synergy Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces, Specify, and Vec

    P840001/S614 · PMA supplement
  82. Altaviva

    P240011/S003 · PMA supplement
  83. InterStim Therapy System, Verify Evaluation System

    P970004/S439 · PMA supplement
  84. Master Restore, Itrel, Synergy, Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces, Specify, and Ve

    P840001/S613 · PMA supplement
  85. InterStim Therapy System, Verify Evaluation System

    P080025/S334 · PMA supplement
  86. Master Restore, Itrel, Synergy, Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces, Specify

    P840001/S610 · PMA supplement
  87. InterStim Therapy System, Verify Evaluation System

    P970004/S426 · PMA supplement
  88. Master Restore, Itrel, Synergy Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces, Specify,

    P840001/S596 · PMA supplement
  89. nterStim Therapy System, Verify Evaluation System

    P080025/S321 · PMA supplement
  90. Master Restore, Itrel, Synergy, Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces-Quad, Spe

    P840001/S612 · PMA supplement
  91. Altaviva Implantable Neurostimulator

    P240011/S002 · PMA supplement
  92. Altaviva Implantable Neurostimulator

    P240011/S001 · PMA supplement
  93. InterStim Therapy System, Verify Evaluation System

    P970004/S438 · PMA supplement
  94. Restore, Itrel, Synergy IntellisTM, VantaTM and InceptivTM Spinal Cord Stimulation Systems and PiscesTM, Specify, and

    P840001/S611 · PMA supplement
  95. InterStim Therapy System, Verify Evaluation System

    P080025/S333 · PMA supplement
  96. Medtronic Altaviva System

    P240011 · PMA decision
  97. InterStim Therapy System, Verify Evaluation System (SNS Urinary Neurostimulator 97800); (SNS Urinary Neurostimulator I

    P970004/S435 · PMA supplement
  98. InterStim Therapy System, Verify Evaluation System (SNS Bowel Neurostimulator 97800); (SNS Bowel Neurostimulators Impl

    P080025/S330 · PMA supplement
  99. Master Restore, Itrel, Synergy, Intellis and Vanta Spinal Cord Stimulation Systems and Pisces, Specify, and Vectri

    P840001/S606 · PMA supplement
  100. Master Restore, Itrel, Synergy IntellisTM, VantaTM and InceptivTM Spinal Cord Stimulation Systems and PiscesTM, Specify

    P840001/S607 · PMA supplement
  101. InterStim Therapy System, Verify Evaluation System

    P080025/S331 · PMA supplement
  102. InterStim Therapy System, Verify Evaluation System

    P970004/S436 · PMA supplement
  103. InterStim Therapy System

    P970004/S421 · PMA supplement
  104. InterStim Therapy System

    P080025/S316 · PMA supplement
  105. InterStim Therapy System; Verify Evaluation System

    P970004/S434 · PMA supplement
  106. InterStim Therapy System; Verify Evaluation System

    P080025/S329 · PMA supplement
  107. Master Restore, Itrel, Synergy IntellisTM, VantaTM and InceptivTM Spinal Cord Stimulation Systems and PiscesTM, Specify

    P840001/S605 · PMA supplement
  108. InterStim Therapy System; Verify Evaluation System

    P080025/S328 · PMA supplement
  109. InterStim Therapy System; Verify Evaluation System

    P970004/S433 · PMA supplement
  110. Master Restore, Itrel, Synergy, Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces-Quad, Spe

    P840001/S604 · PMA supplement
  111. Master Restore, Itrel, Synergy, Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces, Specify

    P840001/S602 · PMA supplement
  112. Master Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems

    P840001/S603 · PMA supplement
  113. InterStim Therapy System, Verify Evaluation System

    P970004/S432 · PMA supplement
  114. InterStim Therapy System, Verify Evaluation System

    P080025/S327 · PMA supplement
  115. InterStim Therapy System, Verify Evaluation System

    P970004/S431 · PMA supplement
  116. InterStim Therapy System, Verify Evaluation System

    P080025/S326 · PMA supplement
  117. Model A71100 Restore Clinician Programmer Application (CPA) and Model A71200 Vanta/Sequentia LT Clinician Programmer A

    P840001/S594 · PMA supplement
  118. InterStim Therapy System; Verify Evaluation System

    P080025/S325 · PMA supplement
  119. InterStim Therapy System; Verify Evaluation System

    P970004/S430 · PMA supplement
  120. Master Restore, Itrel, Synergy Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces, Specify,

    P840001/S601 · PMA supplement
  121. InterStim Therapy System, Verify Evaluation System

    P080025/S324 · PMA supplement
  122. Master Restore, Itrel, Synergy, Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces-Quad, Spe

    P840001/S599 · PMA supplement
  123. InterStim Therapy System, Verify Evaluation System

    P970004/S427 · PMA supplement
  124. InterStim Therapy System, Verify Evaluation System

    P080025/S322 · PMA supplement
  125. InterStim Therapy System, Verify Evaluation System

    P970004/S428 · PMA supplement
  126. Master Restore, Itrel, Synergy Intellis, Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces, Specify, a

    P840001/S598 · PMA supplement
  127. Restore; Itrel; Synergy Intellis; Vanta and Inceptiv Spinal Cord Stimulation Systems and Pisces; Specify; and Vectris Sp

    P840001/S600 · PMA supplement
  128. MEDTRONIC INTERSTIM SACRAL NERVE STIMULATION THERAPY SYSTEM

    P080025 · PMA decision
  129. MEDTRONIC INTERSTIM THERAPY SYSTEM FOR URINARY CONTROL

    P970004 · PMA decision
  130. ITREL(R) TOTALLY IMPLANTABLE SPINAL CORD STIM. SYS

    P840001 · PMA decision
Source information

Source updated Nov 9, 2026. Retrieved Nov 21, 2026.