THINK Surgical, Inc. appears in FDA device records as the applicant for TMINI Miniature Robotic System submissions under product code OLO and regulation 21 CFR 882.4560, the Class II stereotaxic instrument category. The supplied collection links one 510(k) record, K263234, decided Substantially Equivalent on October 16, 2026, and one device recall entry, Z-0192-2027, concerning a TMINI square driver component. Company-published material describes the TMINI Robot as a miniature, handheld, wireless robotic system intended to enable accurate and precise implant placement, with CT-based 3D surgical planning and overhead active tracking. [1][2][5]
Documented product and regulatory identity
The TMINI Miniature Robotic System is filed under regulation number 21 CFR 882.4560, regulation name Stereotaxic Instrument, regulatory Class II, product code OLO. The K263234 record lists THINK Surgical, Inc. as applicant at 47201 Lakeview Blvd., Fremont, California, with a decision code of SESE and decision description Substantially Equivalent, dated October 16, 2026. The same record identifies the review advisory committee as Orthopedic and the advisory committee as Neurology. [1][3]
THINK Surgical's own product page describes the TMINI Robot as a miniature, handheld, wireless robotic system designed to enable accurate and precise implant placement, integrating into the operating room with overhead active tracking and unrestricted patient access. The company states that CT-based 3D surgical planning and the TMINI PRO Workflow enable intraoperative adjustments to help fine-tune implant positioning and stability, and that the system is an open implant robotic system intended to support surgeon implant choice. [5][6]
Submission history in the supplied filings
The linked portfolio record for this collection is K263234, received September 16, 2026 and decided October 16, 2026, with a summary statement and a clearance type listed as Special. The collection counts one 510(k) record with a SESE decision; the sample decision counts likewise show one SESE entry, reflecting the same single linked record rather than a broader portfolio total. [1]
Other TMINI submissions appear in the numbered sources as separate clearances. K250877, dated June 20, 2025, covers software modifications and compares the device to the TMINI Miniature Robotic System cleared via K243481. K252544, dated October 10, 2025, addresses the TMINI Miniature Robotic System (AIM 3.0) against the predicate cleared via K250877. K260010, dated March 3, 2026, concerns the TMINI Miniature Robotic System (AIM 3-2) and lists K253661 as the subject device number in its comparison table. [7][8][9]
Recall record for a system component
Recall Z-0192-2027 is a Class II, firm-initiated voluntary recall of the Think Surgical Square Driver, REF 109126, lot 25356, covering 15 units distributed in New York and Delaware. FDA records give a recall initiation date of August 20, 2026 and a center classification date of October 16, 2026, with the status listed as ongoing. The stated reason is that the robotic system square drive failed to meet design specifications, potentially resulting in a pin becoming stuck in the square driver or the pin wobbling during drilling. [2][4]
Applicant status and what the record does not establish
The FDA records identify THINK Surgical, Inc. as applicant and recalling firm, with a Fremont, California address used in filings; that filing address is not evidence of a headquarters designation. No parent company, brand ownership, customer list, certification or commercial availability claim is established by the public documents. Company descriptions of accuracy, precision and workflow capability originate with THINK Surgical's published material, and the brochure notes that a named physician consultant is a paid consultant of the company. [2][5][6]