Healthmark Industries Co., Inc. appears in FDA device enforcement records as the recalling firm for three 2026 Class II actions, all listed with a Fraser, Michigan address in the filing data. The records cover two product families: vinyl vented instrument guards sold under multiple model numbers, and ProChek protein test products used in cleaning verification. The company’s own published product documentation describes vented instrument guards as gas- or steam-permeable accessories intended to protect surgical instruments through handling and sterilization. The filings describe the scope of each action, the stated reasons and the distribution patterns recorded by FDA. [1][2][3][5]
Vinyl instrument guards: labeling action
Recall Z-0188-2027 covers vinyl instrument guards listed under model numbers 09-1-011 through 09-1-020 and 09-2-011, spanning vented guards in white, blue, green, red, yellow, maroon, orange, brown, black, opaque assorted and transparent blue. The FDA record lists lot numbers tied to specific model numbers and a UDI DI for at least one model. The stated reason is labeling not compliant with 21 CFR 801, with updated steam sterilization instructions. The action was initiated September 7, 2026, and classified October 15, 2026, with status ongoing. [1][6]
The distribution pattern recorded for this action is worldwide, covering the United States nationwide plus Japan, Australia, Canada, Portugal, France, Poland, Puerto Rico, Saudi Arabia, Hong Kong, Bahrain, Qatar, Oman, Israel, Mexico, the United Arab Emirates, Singapore, Bermuda, South Korea, Thailand and the United Kingdom. FDA lists the notification approach as two or more of email, fax, letter, press release, telephone or visit. A third-party recall summary describes the action as voluntary and rated Class II, moderate risk. [1][10]
ProChek protein tests: expired ingredient
Recall Z-3031-2026 covers healthmark ProChek-II under model numbers PT-313 and PT-201-12, with serial and lot numbers enumerated in the FDA record and a stated quantity of 5,090 units. Recall Z-3032-2026 covers healthmark ProChek-W Protein Test for Robotic Arms and ProChek-Semi Quantitative Protein Test, model PT-SQ-001, with listed serial and lot numbers and a stated quantity of 53 units. Both are Class II, voluntary firm-initiated actions sharing event ID 99537. [2][3]
The stated reason for both ProChek actions is a test manufactured with an expired ingredient that may present as either a false negative or positive. Both were initiated July 29, 2026, and classified August 26, 2026, with status ongoing. Distribution for both is recorded as US nationwide plus Canada, France, Hong Kong and South Korea. An FDA recall database page for the ProChek action lists the product classification as qualitative cleaning process protein and shows the record last updated October 30, 2026. [2][3][4]
Published product documentation
Healthmark’s published Vented Instrument Guards page describes the guards as intended to protect surgical instruments through handling and sterilization, gas or steam permeable, suitable for flash and conventional autoclave processes, with solid colors radiopaque for visibility, made without natural rubber latex and non-sterile. The same page lists a price of $1.00 and an inquiry option. A separate Healthmark instrument care document lists item numbers and dimensions for vented and duo-vented guards, including 09-1-014 and 09-1-015. [5][8]
The instrument care document also carries company contact details at 18600 Malyn Blvd., Fraser, Michigan, with toll-free, international, fax and email contacts. That address matches the address field in the FDA enforcement records for the three 2026 actions. The published documentation does not state ownership, parent company or corporate history. [5][8]
What the records establish
The numbered sources establish three 2026 Class II recalls by this firm, their product identifiers, stated reasons, quantities where given, initiation and classification dates, and distribution patterns. They do not establish total company sales, total product portfolio, safety performance across products, or clinical performance of any test or guard. The FDA.report entry for Z-0188-2027 lists product code MDM, firm FEI 1000160387 and root cause as labeling design, with a set of 510(k) numbers; those clearances are not described here beyond their listing. [1][2][3][6]